J. B. MOHLER
Director of Research, Johnson Bronze Company, Newcastle, Pa.
INTRODUCTION
Plating-range cells have become widely
accepted as laboratory tools for electroplating control and experimentation.
Calibrations were reported by Hull(1)
for the Hull cell, the best known and most popular plating-range cell. These
calibrations were made by stripping and weighing small sections of the deposits
and by measuring the current density at points along the cathode with a current-density
meter. An equation for general use was derived by averaging data for nickel,
acid copper, cyanide zinc and cyanide cadmium solutions, but it was pointed
out that individual calibrations are necessary for accurate interpretations
of results.
By the use of thickness measurements
a cell may be calibrated not only for plate thickness, but where the cathode
efficiency is 100 per cent or where the variation in cathode efficiency with
current density has been measured, also for current density.
Measurements carried out under project
No. 7 of the American Electroplaters Society(2) show the relative accuracy
of various methods for measuring the thickness of metal deposits. These studies
may be used as a guide in the calibration of plating-range cathodes. If a rapid
calibration is required the Magne-gage is a very convenient instrument.
Many factors can influence the metal
distribution over a cathode, as shown by Tucker and Flint(3). Not only does
cathode efficiency and covering power vary with the preparation of the test
panel, but in some baths, such as the alkaline tin bath, it also varies with
the thickness of the deposit when thin layers are concerned. Therefore, if a
plating-range cell is to be used for precise work it is sometimes necessary
to calibrate it under the expected conditions of use.
For those baths that have a cathode
efficiency of 100 per cent, however, it is likely that the calibration of a
plating-range cell for a particular bath will apply under various conditions
of time and total current. For that reason, in the work reported here, it was
decided to use two such baths in the calibration of a slot cell(4) with a slot
at one end as shown in Fig. 1.
THE SLOT CELL
There are two distinct advantages to this cell. First, the anode is outside
of the cell and thus has no influence on the cathode. Second, the cell may be
immersed in a plating bath to overcome the difficulties encountered with temperature
control in box-type plating range cells.
Another useful feature of the cell
is that it is dimensionless. The cell shown in Fig. 1 is 4 inches (10.2 cm)
long with a 1/4-inch (6.4-mm) slot 1 inch (25.4 mm) from the cathode. A 2-inch
(5.1-cm) cell was constructed with a 1/8-inch (3.2-mm) slot inch (12.7 mm) from
the cathode Test cathodes were plated in these cells using an acid tin bath
at an average cathode current density of 10 asf (1.1 amp/dm2). For a 5-minute
plating time the plate pattern in each case agreed very closely with that shown
at the top of Fig. 2 when the slot-cathode distance was used as the unit for
measuring the distance on the panel from the slot end. For long plating times
this relation would not hold, because the ratio of cathode area to cell volume
changes with the size of the cell, and the solution in a small cell would become
depleted of metal more rapidly than the solution in a large cell.
Fig 1. Slot type plating
range cell; at top, seen from front (dashed horizontal line indicating solution
level) and at bottom, in cross section
CALIBRATION OF A SLOT CELL
Two baths were used to calibrate the 4-inch slot cell, a silver cyanide bath
and an acid tin bath. These baths were chosen because they both deposit metal
at essentially 100 per cent cathode efficiency and because they are representative
of baths of relatively high and relatively low throwing power, respectively.
Fig. 2. At top, appearance
of plate, and at bottom, current distribution, on panel plated in slot-type
plating-range cell immersed in an acid tin solution
The acid tin bath had the following
composition:
g/l
Tin
50
Free fluoboric acid
40
Boric acid
20
Cresol sulfonic acid
30
Beta naphthol
1
Gelatine
1
A 2 1/2 x 4-inch (6.4 x 10.2-cm)
polished steel cathode was used with 2 inches (5.1 cm) immersed in the bath.
A current of 0.56 amperes was allowed to flow for 40 minutes. The plate was
then stripped from the cathode, and vertical sections were cut from the tin
sheet. These were weighed, and from the weight the current densities at points
along the cathode were calculated. Above a current density of 35 asf (3.8 amp/dm2)
the deposit was quite rough and was not used.
Three panels of the same size were
cleaned, pickled and plated in the same manner, but for minutes. Magne-gage
readings were taken on the three panels and averaged. From the thickness determinations
the average current densities were calculated.
Fig. 3. At top, appearance
of plate, and at bottom, current distribution, on panel plated in slot-type
plating-range cell immersed in cyanide silver solution
The current densities obtained by
the two methods are shown in Fig. 2.
The silver bath had the following
composition:
g/l
Silver nitrate
60
Free potassium cyanide
50
Potassium hydroxide
5
A 2 1/2 x 4-inch (6.4 x 10.2 cm) polished steel cathode was used with 2 inches
(5.1 cm) immersed in the bath. A current of 0.22 amperes was allowed to flow
for 1 hour. The plate was then stripped from the cathode, and vertical sections
1/8 inch (3.2 mm) wide by 1 inch (25.4 mm) long were cut from the silver sheet.
These were weighed, and from the weight the current densities at points along
the cathode were calculated.
Another cell was constructed, of-
the same proportions as the 4-inch (10.2-cm) cell but of three times its size,
i. e., the slot was inch (1.9 cm) wide and placed 3 inches (.6 cm) from a 12-inch
(30.5-cm) long cathode. The depth of immersion during plating was 2 inches (5.1
cm). Three cathodes were cleaned, pickled and struck with silver prior to plating
for 10 minutes at 0.67 amperes. Magne-gage readings were taken, and the current
densities were calculated from the averages for the three cathodes.
Fig. 3 shows the two sets of data
for the silver bath plotted on the same graph where the distances from the high
current density end are in terms of the slot-to-cathode distance as the unit.
The data show that a rapid calibration
can be made with the Magne-gage, and that the calibration for a cell of one
size can be used for another of equal proportions.
There is an apparent difference in
the shape of; the curves obtained by the two methods of calibration, particularly
for the silver solution at low current densities. An investigation to determine
whether this was due to the difference between the two methods of thickness
measurement or to differences between the bonded and unbonded deposits would
have required considerable more experimentation.
CURRENT DENSITY DISTRIBUTION EQUATIONS
Approximating equations were derived by plotting the logarithm of the average
current -density against the distance from the slot end and solving for the
point-slope formula. These equations are as follows:
For the acid tin solution, c = antilog
0.50 (1.84 - z)
For the silver solution, c = antilog
0.36 (1.78 - x)
where c = Times average current density
x = Distance along cathode in slot-cathode-distance
units
Equations were also derived by using
the increase in thickness as one moves 1 unit of distance from any point as
a denominator power function and the ratio of the theoretical thickness at zero
distance to actual thickness at the point of average current density as the
numerator. These equations are as follows:
For the acid tin solution,
8.1
c
=
3.2x
For the silver solution,
4.1
c
=
2.2x
Fig. 4. Current distribution
curves obtained in slot-type plating-range cells. Note the more uniform
current distribution on panels plated in solutions of high throwing power
The curves represented by these equations
follow very closely the solid-line curves in the graphs.
The equations did not hold near the
high-current density end of the panel, because there the current density tended
to level off, and the results were erratic. Whether or not the other end of
the cell would interfere with the shape of the curve was not determined, but
it is suspected that it does; this difficulty can be overcome, however, merely
by using a cell that is longer than the distance under study.
CURRENT DISTRIBUTION VS. THROWING
POWER
Most of the curve obtained in this type of cell approximates a simple exponential
curve. With more accurate calibrations it may be possible to obtain an indication
of the effect of polarization over the entire range, or in more popular terms,
to obtain an indication of change in throwing power with change in current density
over the entire range. The curves in Fig. 4 were obtained from data from various
sources and are only approximate. They do, however, provide a good graphical
illustration of the effect of throwing power.
APPLICATION OF THE CELLS
A cell that can be used to deposit metal of a thickness that changes continuously
in a known manner is a useful laboratory tool. In addition to its use as a plating-range
cell in which the behavior of a bath can be studied, it may be employed where
variable plate thickness on a single cathode offers an advantage. For example,
it was desired to determine the minimum thickness of silver required to prevent
rust spots, during air drying following a hot rinse, on steel ground to 100-microinch
(2.5-µ) roughness. By processing blanks in a slot-type cell this thickness
was readily determined, visually, to be 0.3 mil (8 µ). Similar obvious
applications would be found in studies of porosity and corrosion.
REFERENCES CITED
(1). R. O. Hull, Proc. Am. Electroplaters Soc. 27, 52 (1939)
(2). H. J. Read and F. R. Lorenz, Plating 38, 255 (March, 1951) and 945 (September,
1951).
(3). W. M. Tucker and R. L. Flint Trans. Electrochem. Soc. 88, 338 (1945).
(4). J. B. Mohler and R. A. Schaefer, Monthly Rev. Am. Electroplaters
Soc. 34, 1361 (December, 1947).
The information contained in this site is provided for your review and convenience. It is not intended to provide legal advice with respect to any federal, state, or local regulation.
You should consult with legal counsel and appropriate authorities before interpreting any regulations or undertaking any specific course of action.
Please note that many of the regulatory discussions on STERC refer to federal regulations. In many cases, states or local governments have promulgated relevant rules and standards
that are different and/or more stringent than the federal regulations. Therefore, to assure full compliance, you should investigate and comply with all applicable federal, state and local regulations.